4-DIGIT OHMMETER, TRANSMITTERS & CONTROLLERS MODEL Q9000O
OPTIONS Analog output for user-selected span Single-setpoint 10 A relay Dual-setpoint 10 A relays Proportional 4-20 mA control Time-proportional 2 A SS relay control Isolated parallel BCD output NEMA-4 splash-proof lens cover DISPLAY AND POWER OPTIONS The QUANTA Model Q9000O is available with an LED or an LCD display and with six types of meter power: 120 V ac, 240 V ac, 24 V ac, 5 V dc, 9-32 V dc (isolated) and 26-56 V dc (isolated). ANALOG OUTPUT OPTIONS A 0.1 mV/count (±2 V full-scale) analog output is standard and is ideal for driving a strip-chart recorder. An additional analog output can be provided by an optional vertical plug-in board. Available output signals are 0-5 V dc, 0-10 V dc, 0-1 mA (source or sink), and 4-20 mA (source or sink). The top and bottom of each output range can be scaled to fit a user-selected display span. CONTROL OUTPUT OPTIONS Additional outputs can be provided by a horizontal upper board. Available options include single-setpoint control with one 10 A relay, dual-setpoint control with two 10 A relays, 4-20 mA proportional control (source or sink), time- proportional 2 A solid-state relay control, and isolated, parallel BCD output. The 10A relay options are ideal for ATE and material testing applications, which require ON/OFF control. The BCD output can communicate with programmable controllers and Newport's Model 872A thumbwheel controller. For additional information, please refer to the QUANTA and Mechanical sections. SPECIFICATIONS INPUT
INPUT CONFIGURATION Connection: 2- or 4-wire Zero adjustment: ±50 counts Overvoltage protection (differential): 15 Vp Open-circuit voltage (max): 12 V NOISE REJECTION NMR: 115dB, 50/60 Hz CMR: 120 dB CMV: 1500 Vp per HV test, 354 Vp per IEC spacing ACCURACY AT 25°C Overall accuracy: ±0.05% of reading ±1 count Zero tempco: ±2.5 mohm/°C ±0.001% of FS/°C Span tempco: ±0.006% of reading/°C Warmup to rated accuracy: 1 min ANALOG-TO-DIGITAL CONVERSION Input configuration: Differential, bipolar Technique: Dual-slope, average-value Polarity: Automatic Signal integration period: 100 ms Read rate: 2.5/s DISPLAY Type: 7-segment, red LED Height: 0.56 in (14.2 mm) Symbols: 8.8.8.8 Decimal-point position: Three positions selectable by jumpers behind lens or at connector 0 to 9,999 counts: Normal operation 10,000 to 19,999 counts: Four least-significant digits POWER AC voltages: 120, 240, 100 or 24 V ac +10%/15% AC frequency: 49-440 Hz DC voltages: 9-32 V dc, isolated to 300 Vp; 26-56 V dc, isolated to 300 Vp; 5 V dc ±5%, non-isolated DC power consumption: 5 W max ENVIRONMENTAL Operating temperature: 0 to 60°C Storage temperature: 40 to +85°C Relative humidity: 95% at 40°C (non-condensing) MECHANICAL Dimensions: Newport DIN1A (1/8 DIN) case (See Mechanical section for drawings) Weight: 17 oz (480 g) Case material: 94 V-0 UL-rated polycarbonate HOME / PRODUCT GUIDE / TECHNICAL REFERENCE NEWPORT Electronics, Inc. (714) 540-4914 or Contact Us Copyright 1998, Newport Electronics, Inc. All rights reserved. |